BulletPoint1 : TB100 Turbidimeter Turbidity Meter Continuous measurement mode can be used for indexing or matching the sample vials. Auto-Power Off effectively conserves battery life.
BulletPoint2 : Setup menu allows setting the 7 parameters, including the number of calibration points, resolution, date and time, etc. Reset feature automatically resumes all settings back to factory default options.
BulletPoint3 : Selectable 4 turbidity units, including the NTU, FNU, EBC and ASBC. Single measurement mode automatically recognizes and holds the stable reading.
BulletPoint4 : Expanded memory stores and recalls up to 100 data sets. Stored data can be transferred into computer by USB communication interface. Multi-mode power scheme (battery, power adapter or USB port) ensures that use the meter smoothly.
BulletPoint5 : Easy-to-use portable turbidity meter comes with a standard carrying case. 2 to 5 points push-button calibration using the Formazin Standards.
Color : Black
ExternallyAssignedProductIdentifier1 : 0783761046561
ExternallyAssignedProductIdentifier2 : 783761046561
ItemName : Tongbao Portable TB100 Turbidimeter Turbidity Meter Tester Analyzer 0 to 1100NTU Measurement USB Data 2 to 5 Points Calibration Selectable 4 Turbidity Units NTU FNU EBC and ASBC 100 Data Sets Storage
ItemPackageDimensions_Height : 14 centimeters
ItemPackageDimensions_Length : 45 centimeters
ItemPackageDimensions_Width : 35 centimeters
ItemTypeKeyword : science-lab-turbidity-meters
Manufacturer : Tongbao
Material : Plastic
PartNumber : TB100
PowerSourceType : Battery Powered
ProductDescription :
Introduction:
High performance portable turbidity meter, 2 to 5 points calibration, selectable 4 turbidity units. The meter is suitable for process control and field use.
Unpacking Before unpacking, ensure that the current work environment meet following conditions.
Relative humidity is less than 80%
Ambient temperature is greater than 0C and less than 60C
No potential electromagnetic and ambient light interference
Operating Principles
TB100 portable turbidimeter operates on the nephelometric principle of turbidity measurement and is designed to meet the criteria specified in ISO 7027.
The following figure shows basic optical system of the meter. It includes a light source and a detector to monitor the light scattered at 90 degree with respect to the incident beam.
Specifications
Model : TB100
Measurement Method :ISO 7027 nephelometric method (90°)
Measurement Range : 0~1100 NTU/FNU, 0~275 EBC, 0~9999 ASBC
Resolution : 0.01 (0~100 NTU), 0.1 (100~999 NTU), 1 (999~1100 NTU)
Accuracy :± 2% (0~500 NTU) ± 3% (501~1100 NTU)
Calibration Standards : :0.02 10.00 200 500 1000 NTU
Light Source :Infrared-emitting diode (850 nm wavelengt)
Detector :Silicon :Photodiode
Stray Light : <0.02 NTU
Sample Vials : 60(H)× 25(Dia)mm
Measurement Modes:Single or Continuous measurement
Power Off :Manual or Automatic (15 minutes after last key pressed)
Reset Function:Yes
Memory :Stores up to 100 data sets
Output USB Communication Interface
Operating Temperature : 0~60°C
Power Requirements : 1× 9V battery(battery not included)
Dimensions :180(L)× 85(W)× 70(H)mm
Weight :300g